Bedside checklist: Chapter 8. Fluid Responsiveness and Fluid Tolerance¶ 0 / 39Clear all Indication I stated the clinical question before scanning: is this patient preload responsive, congested, or bothStandard: Question named in the note; study labeled FoCUS or advanced CCESource: 14, 89 I confirmed acute circulatory failure is present before testing responsivenessStandard: Hypotension, low cardiac output or tissue hypoperfusion documented; no test in a stable patientSource: 1, 2 I checked whether the answer is already obvious and testing would only delay treatmentStandard: Hemorrhage, overt hypovolemia or first-phase septic shock before any volume: give fluid, do not testSource: 1, 2 I recorded ventilation mode, tidal volume, PEEP, spontaneous effort, rhythm and intra-abdominal pressure concerns before choosing an indexStandard: Documented, because these determine which index is validSource: 1, 2 I checked contraindications to the test selectedStandard: No PLR in intracranial hypertension; no occlusion test in a non-intubated patient or one who breaks the holdSource: 1, 2 I asked the tolerance question separately from the responsiveness questionStandard: Congestion, oxygenation and organ risk weighed before any bolus is givenSource: 17, 24 I asked whether the right ventricle is the cause of the congestionStandard: Right ventricular size, function and afterload assessed; pericardial constraint excludedSource: 19, 83 I checked whether this patient falls in a category in which one of the venous signals is uninterpretableStandard: Cirrhosis or steatosis for the portal vein, parenchymal renal disease for the intrarenal veins, severe tricuspid regurgitation for the hepatic vein, raised intra-abdominal pressure for the IVC, atrial fibrillation for phase assignmentSource: 17, 19 I declared an abdominal venous study under the ASE nomenclature as a cardiac examination with an examination of another system, and archived it so the noncardiac elements are identifiedStandard: Named as cardiac POCUS with abdominal venous Doppler, not as a cardiac studySource: 20 Acquisition I obtained an LVOT pulsed-wave signal with the closing click visible, low wall filter and gain set to the modal envelopeStandard: 5 to 7 mm gate in the LVOT; outer edge of modal velocity tracedSource: 32 I averaged an adequate number of beatsStandard: Three in sinus rhythm, five in atrial fibrillation within one examination; five to seven consecutive beats when arrhythmia is present during PLRSource: 32, 33 I kept the probe on the chest and the sample volume in the same LVOT position throughout the maneuverStandard: No re-acquisition of the window between baseline and testSource: 2 I performed PLR by moving the bed from semi-recumbent to trunk horizontal with legs at 30 to 45 degreesStandard: Trunk-to-leg angle unchanged; legs not lifted by hand; secretions suctioned and awake patient informed firstSource: 28, 32 I measured the PLR effect within about one minute and again after returning the patient to semi-recumbentStandard: Three paired VTI values recorded: baseline, PLR, returnSource: 28, 32 I imaged the IVC in a subcostal long axis, perpendicular to its long axis, at the site the book measures toStandard: The house measurement of section 6.4, which is 0.5 to 3.0 cm proximal to the RA ostium read at end-expiration; the 1.0 to 2.0 cm from the junction quoted from the 2015 guideline falls inside it; away from the hepatic vein confluence and diaphragmSource: 12, 13, 32 I confirmed on 2D that apparent IVC collapse is narrowing and not translation out of planeStandard: Pseudocollapse excluded before any number is reportedSource: 13, 14 Where the transthoracic answer was equivocal in a ventilated patient, I considered TEE for SVC collapsibilityStandard: SVC measurable in 99.6% of ventilated patients versus 78.1% for IVCSource: 9 I acquired the hepatic vein trace to the standard of section 6.4, with a simultaneous electrocardiogram running so the phases could be assignedStandard: Subcostal, 3 to 4 mm sample volume, gate 1 to 2 cm from the IVC junction, at least five consecutive beats spanning one respiratory cycle when ventilated; ECG channel present on the archived image, since reader agreement is higher with itSource: 13, 54 I set the portal velocity scale and the angle so that both peak and trough velocities were on screen and measurableStandard: Cardiac preset scale reduced from a default near 100 cm/s to about 40 cm/s; sweep 50 or 66.7 mm/s; no aliasing and no falsely absent flowSource: 19 I obtained an intrarenal trace with arterial flow above and venous flow below the baseline for at least two cycles, and rejected an ambiguous study instead of interpreting itStandard: Abdominal or trauma preset, Nyquist limit below 20 cm/s, color gain increased; inadequate trace recorded as uninterpretableSource: 17, 19 For a venous congestion study I measured the IVC at the site the convention specifies, and said which convention I usedStandard: Congestion studies: intra-hepatic portion, 2 cm from the hepatic vein junction, maximal diameter across the respiratory cycle. Right atrial pressure: end-expiration, 0.5 to 3.0 cm proximal to the right atrial ostium, the house measurement of section 6.4Source: 13, 16 Interpretation I applied the threshold belonging to the measurement I actually madeStandard: PLR: cardiac output rise 10 ± 2%, or LVOT VTI rise >12.5%; never a pulse pressure threshold applied to VTISource: 11, 14 I compared the observed change with the least significant change of my own measurementStandard: LVOT VTI LSC 11% same operator, 14% between operators; ASE calls >6% meaningfulSource: 14, 33 I did not report an IVC index as a fluid-responsiveness test with spontaneous effort, low tidal volume, low compliance, intra-abdominal hypertension or right ventricular failureStandard: Index withheld or explicitly flagged as uninterpretableSource: 2, 32 I did not estimate RAP from IVC collapse on positive-pressure ventilationStandard: Reported collapse as uninterpretable; used IVC ≤2.1 cm to identify RAP below 10 mm Hg, hepatic vein velocities over at least five beats, or a transduced invasive pressureSource: 12, 13 I confirmed return to baseline before calling a PLR positiveStandard: Cardiac output returned to the pre-test value in the semi-recumbent positionSource: 28 I computed the portal pulsatility fraction from measured velocitiesStandard: (Vmax - Vmin)/Vmax; below 30% normal, 30% to 50% a transmitted pulse, 50% or more markedSource: 19 I named the false positive and the false negative for each venous bed before accepting itStandard: Portal: lean or athletic subjects and arteriovenous connections as false positives, cirrhosis or steatosis as false negative. Hepatic: severe tricuspid regurgitation and atrial fibrillation as false positives, preserved right atrial compliance as false negativeSource: 16, 17, 19 I read the venous beds as pressure transmitted into the organs, in context, and did not read a plethoric IVC as congestion on its ownStandard: Dilatation alone has 41% specificity for congestive acute kidney injury and can be normal in athletes, young adults and pregnancy; marked findings in at least two beds are 96% specific and 27% sensitive for subsequent injury after cardiac surgerySource: 16 Medical decision-making I gave fluid only when circulatory failure, preload responsiveness and acceptable overload risk coexistedStandard: All three documented; otherwise fluid withheldSource: 24, 28 When the test was negative I named the alternative interventionStandard: Vasopressor, inotrope, PEEP reduction or decongestion chosen instead of a bolusSource: 2, 28 I verified the effect of any bolus given, without assuming it workedStandard: Post-bolus stroke distance and perfusion markers (mottling, capillary refill time, lactate, venous oxygen saturation) recheckedSource: 2, 22 I repeated the test before the next fluid decision, without relying on an earlier resultStandard: In FRESH only 42.2% were responsive on the first test but 81.9% at some point over 72 hoursSource: 23 I assessed preload responsiveness before starting fluid removal or ultrafiltrationStandard: Absence of responsiveness supports safe removal; a PLR before removal predicts intradialytic hypotensionSource: 1, 2 I documented the index used, its threshold, the measurement technique and the ventilatory and rhythm conditions at the timeStandard: Report allows the next operator to repeat the identical measurement; quantitative monitoring requires level II training, level III for Doppler hemodynamicsSource: 14 I said what the venous finding changes, or that it changes nothing, and recorded the standing of the evidence behind itStandard: One of stop point, decongestion target, or nothing, written in the note, with the association to organ injury described as observational, strongest after cardiac surgery, and with no randomized trial of an ICU strategy guided by these signalsSource: 17, 75, 76 I did not treat a venous finding as an indication for fluid removal in the absence of a clinical decongestion targetStandard: No diuresis or ultrafiltration ordered on a venous finding alone; caution documented where cardiac output may depend on high preloadSource: 19 I named the waveform that must change, and by how much, before the next assessment counts as differentStandard: Target stated; no published least significant change exists for the hepatic venous systolic filling fraction or the portal pulsatility fraction, and a change greater than 10% is the rule quoted for the IVC collapsibility indexSource: 14 I reported the estimated right atrial pressure and, where warranted, the hepatic vein waveformStandard: Right atrial pressure reported, even as "unable to assess"; hepatic vein waveform reported for volume status among other indicationsSource: 21 Download the PDF · Back to the chapter